 Continuously monitor the level of liquids and slurries, without worry of false signals as a result of material build-up. Loop powered Series CRF delivers a 4-20 mA isolated output signal proportional to the height of material in a tank or holding vessel. Continuous level transmitter can be field calibrated to fit your specific application. Integral electronics and two-wire design simplify installation. Probes should be mounted vertically and are easily trimmed to shorter lengths where required. When mounting in nonmetallic or horizontal tanks, a separate grounding rod must be used. Consult factory regarding mounting flange and grounding rod assemblies.
PHYSICAL DATA
| Supply Voltage: |
|
18 to 30 VDC |
| Output Signal: |
|
4-20 mA, 2-wire, isolated |
| Maximum Loop Resistance: |
|
600W @ 24 VDC |
| Zero Range: |
|
0 to 500 pF |
| Repeatability: |
|
0.1% FS |
| Operating Temperature Range: |
|
-10 to 180°F (-23 to 82°C) |
| Mounting: |
|
¾" NPT(M), Optional flange mounting |
| Conduit Opening: |
|
½" NPT (F) |
| Probe Enclosure: |
|
NEMA 4, 7 & 9, epoxy coated aluminum |
| Wetted Materials: |
|
316SS, Teflon® |
| Power Consumption: |
|
6VA |
| Span Range: |
|
25 to 2200 pF |
| Linearity: |
|
0.3% of range |
| Weight: |
|
7.25 lbs. (3.29 kg) |
APPLICATIONS
Alcohols, wastewater, oils, polymers, acids, caustics, pulp, slurries, wood chips, water-based products in tanks, bins, or other vessels.
|